Operating Pressure Range +/- 0.250 inches Water Column ( +/- 6.35 mm WC, +/- 62.5 Pascals)
Accuracy +/- 0.003 inches Water Column or +/- 1% of reading whichever is greater
Resolution 0.001 inches Water Column ( +/- 0.05 mm WC, +/- 0.5 Pascals)
Burst Pressure 3 psi (20 kPa) on either inlet nozzle
Pressure Units "WC (inches Water Column), mmWC (millimeters Water Column) or Pa (Pascals)
Data (Log) Storage 4 MB non-volatile memory
Display 4.3" Color Touchscreen LCD (480 x 272 pixels)
Internal Clock The internal clock is powered by a self-charging lithium-ion battery that provides clock operation when AC power is not present
Printer 20 character wide thermal printer (uses 2.2" wide thermal printer paper)
Pressure Reading Log Rates (Log Rate) Normal Operation -- Highest and lowest pressure readings are logged & printed at intervals of 5, 15, 30 minutes or OFF
Alarm Condition -- Highest and lowest pressure readings are logged & printed at intervals of 15, 30, 60 or 120 seconds for first 10 minutes of alarm condition, increasing to 15 minute intervals thereafter
Alarms Two programmable alarm settings, 95db audible alarm and red bordered on-screen warning indicate alarm condition
Cellular Service Cellular modem can email status changes up to four times per hour and the jog log of the last 24 hours once a day. First two years are free. Service is provide by Verizon and is only available in areas with Verizon coverage in the USA.
WiFi WiFi can email status changes up to four times per hour and the jog log of the last 24 hours once a day. Supports 802.11b/g/n with WEP, WPA or WPA2.
Pressure Inlet Nozzles Two 3/16" OD barbed hose connectors, 10 ft of pressure hose provided
USB Port USB V1.1 Type-A
USB Port USB V1.1 Type-B
AC Power 80-240 VAC 50-60Hz 10W (with 6 ft power cord)
Battery Power Internal rechargeable 3.7 Volt Li-ion type battery provides 4-6 hours of portable operation, recharges in 4 hours when AC power is available (printer is disabled when powered by battery)
Operating Temperature Range 32°F to 104°F (0°C to 40°C)
Case Dimensions 9.25" x 7.5" x 4.5", Shipping Weight 8 lbs., case & handle is copolymer polypropylene with polycarbonate window in the lid, and stainless steel hanging hook
Warranty One Year Limited Warranty